Details
Features
- 2 ADC (20-bit) for PPG acquisition
- 1 ADC (20-bit) for ECG/BIOZ acquisition
- 8 photodiode input pins
- 8 emitters driven by 2 drivers up to 300mA at 500mV
- PPG SNR up to 120dB FS (white card loop back test)
- Advanced ambient light offset cancellation (AAOC)
- Excellent ECG, CMRR ~ 118dB, Noise ~ 0.68µVRMS
- Power consumption down to < 20µA @ 25 SPS
- Package: 2.815 x 2.545 x 0.5 mm³, WLCSP, 42 pins
Benefits
- Lowest noise ECG frontend (0.68µVRMS)
- Synchronized PPG and ECG for blood pressure
- Lowest power consumption on the market
- Innovative functions to measure galvanic skin response and bio-impedance (body composition)
- High variety and number of sensing inputs allow implementation of the widest variety of vital sign monitoring technologies for consumer medical applications and smart watches, fitness bands, monitoring patches
Parameters
Channels
input
8
output
8
Supported algorithms
Heart rate
Heart rate variability
Blood pressure
SpO2
Package
WLCSP
Package dimension
l
2.82
mm
w
2.55
mm
h
0.4
mm
Supply voltage
max.
1.98
V
Interface
input
I²C slave
SPI (16MHz)
output
I²C slave
SPI (16MHz)
Operating ambient temperature
min.
-30
°C
max.
85
°C
SNR
120
dB
Application fields
Ordering codes
Downloads
Datasheets
Short datasheet
Environmental documents
Material declaration form
Material declaration form
Material declaration form
Material declaration form
RoHS certificate
Images
Images
Block diagram
Software
Evaluation software
Evaluation software
Library
Product selector
Find the right product.
More analog frontend
State-of-the-art Analog Front End ICs comprise a one chip solution to derive high quality digital signals.